LCR Meter

Catalogue

  • Hioki LCR Meters and Impedance Analyzers range from 1 mHz to 3 GHz devices to suit a wide range of applications in the testing of electronic components. The IM3523A is designed for basic measurement on production lines so that it can be offered at a low price point, and delivers improved measurement reliability thanks to its contact check function.

Measurement modes

LCR (Measurement with single condition), Continuous testing (Continuous measurement under saved conditions)

Measurement parameters

Z, Y, θ, X, G, B, Q, Rdc (DC resistance), Rs (ESR), Rp, Ls, Lp, Cs, Cp, D (tanδ)

Measurement range

100 mΩ to 100 MΩ, 10 ranges (All parameters defined in terms of Z.)

Basic accuracy

Z : ±0.05% rdg. θ: ±0.03°

Measurement frequency

40 Hz to 200 kHz (5 digits setting resolution)

Displayable range

Z, Y, Rs, Rp, Rdc, X, G, B, Ls, Lp, Cs, Cp : ± (0.00000 [unit] to 9.99999G [unit]) Real value display for Z and Y only θ: ± (0.000° to 180.000°), D: ± (0.00000 to 9.99999) Q: ± (0.00 to 99999.9), Δ%: ± (0.0000% to 999.999%)

Measurement signal level

V mode, CV mode: 5 mV to 5 Vrms, 1 mVrms steps CC mode: 10 μA to 50 mArms, 10 μArms steps

Output impedance

100 Ω

Display

Monochrome LCD

Measurement time

2 ms (1 kHz, FAST, representative value)

Functions

Comparator, BIN measurement (classify function), Panel loading/saving, Memory function

Interfaces

EXT I/O (handler), USB communication (full-speed), LAN

Power supply

100 to 240 V AC, 50/60 Hz, 50 VA max

Dimensions and mass

260 mm (10.24 in) W × 88 mm (3.46 in) H × 203 mm (7.99 in) D, 2.1 kg (84.7 oz)

Included accessories

Power cord × 1 , CD-R (Includes Instruction manual, PC commands and Sample software) × 1, Startup guide× 1, Operating precautions× 1