Number of channels |
12 |
Maximum in-series connections |
In-series connections of instrument up to and including a maximum in-series output voltage of 1000 V |
Voltage output accuracy |
±0.0150% of setting ±500 μV |
Voltage measurement accuracy |
±0.0100% of reading ±100 μV |
Output range |
DC voltage: 0.0000 V to 5.0250 V (set independently for all channels) Maximum output current: ±1.00000 A (set independently for all channels) |
Measurement range |
DC voltage: -0.00100 V to 5.10000 V DC current (2-range architecture): ±1.20000 A (1 A range) ±120.0000 μA (100 μA range) |
Integration time |
1 PLC (50 Hz: 20 ms; 60 Hz: 16.7 ms) × number of smoothing iterations (user-configured) |
Current measurement accuracy |
1 A range: ±0.0700% of reading ±100 μA 100 μA range: ±0.0350% of reading ±10 nA |
Interfaces |
LAN |
Power supply |
Universal (100 V to 240 V AC), 50 Hz / 60 Hz |
Dimensions and mass |
430 (16.93 in)W × 132 (5.20 in)H × 483 (19.02 in)D, 10.3 kg (363.3 oz.) |
Included accessories |
User manual × 1, power cord × 1, rack frame × 1, disk with computer application × 1 (Available within the range of application specifications) |