AC Power Clamp Meter

Catalogue

  • Hioki clamp meters are best in class handheld current and voltage meters for safely and accurately measuring power lines. The CM3286-50 clamp on power meter can measure voltage, current, power, power factor, phase angle, reactive power or frequency, as well as detect phase sequence on live lines at any desired point of a single-phase circuit or three-phase circuit.

Measurement line

Single-phase, Three-phase (balanced with no distortion)

Measurement items

Voltage, Current, Voltage/Current peak, Active/Reactive/Apparent power, Power factor, Phase angle (*1), Frequency, Simple Active Energy Consumption (Single-phase) [With Z3210 installed (*2)] Voltage/Current harmonics

AC voltage range

[Measurement range] 80.0 V to 600.0 V, Single range, Basic accuracy 45-66 Hz: ±0.7% rdg. ±3 dgt. (Frequency characteristics: 45 to 1 kHz, True RMS)

AC current range

[Measurement range] 0.060 A to 600.0 A, 3 range, Basic accuracy 45-66 Hz: ±1.3% rdg. ±3 dgt. (Frequency characteristics: 45 to 1 kHz, True RMS)

Power range

[Single phase] 0.005 kW to 360.0 kW Basic accuracy: ±2.0% rdg. ±7 dgt. (50/ 60 Hz, Power factor=1) [Balanced three-phase 3-wire] 0.020 kW to 623.5 kW Basic accuracy: ±3.0% rdg. ±10 dgt. (50/ 60 Hz, Power factor=1) [Balanced three-phase 4-wire] 0.040 kW to 1080 kW Basic accuracy: ±2.0% rdg. ±3 dgt. (50/ 60 Hz, Power factor=1)

Harmonic levels

[With Z3210 installed (*2)] Voltage/Current harmonic levels up to 30th, Content factor, Total harmonic distortion ratio

Other functions

[Phase angle (*1)] lead -180.0° to lag 179.9°, [Power factor] -1.000 to 1.000 [Frequency] 45.0 Hz to 999.9 Hz, PEAK, Phase detection, MAX/MIN/AVG value display, Auto hold, electric meter comparison, unbalanced 3-phase power estimate display, etc.

Dustproof and waterproof

IP20 (Voltage measurement in a completely dry condition. When jaw closes) IP50 (While in storage)

Power supply

LR03 Alkaline battery ×2 Continuous use: approx. 25 hr (without Z3210 installed), approx. 18 hr. (with Z3210 installed and using wireless communications) Other conditions: 100 A AC measurement, backlight off, 23°C reference value

Core jaw dia.

φ 46 mm (1.81 in), Jaw dimensions: 92 mm (3.62 in) W x 18 mm (0.71 in) D mm

Dimensions and mass

65 mm (2.56in) W × 241 mm (9.49in) H × 35 mm (1.38in) D, 450 g (15.9 oz)

Included accessories

Connection Cord L9257 ×1, LR03 Alkaline battery ×2, Carrying Case C0203 ×1, Instruction Manual ×2, Operating Precautions ×1